Description
DMS 2 Ultrasonic Thickness Gauge
The DMS 2 is a revolutionary A-Scan Digital Thickness Gauge/Data Recorder incorporating beneficial performance features, ease of use, and KrautkramerÂ’s patented (US Pat. 6,035,717) TopCOAT Technology and Auto-V (Auto Velocity) measurement modes.. All pertinent inspection data is displayed to the user on a large high resolution 1/4 VGA display designed to use direct sunlight to enhance the contrast.
The DMS 2 is built to last in a custom-molded, gasket sealed high-impact resistant housing. The robust rubber field case provides added protection for withstanding use in harsh industrial environments.
Specifications
Operating Principle
Ultrasonic pulse/echo
Measurement Modes
Dual-element, Dual-element Dual-multi,
Single-element initial pulse-to-1st, singleelement multiple echo (peak and flank
modes), B-scan (optional TopCOAT & Auto-V
modes)
B-Scan
Cross-sectional view quickly identifies
thinned or laminated areas. Timed 3.5-21
seconds
Probe Zero Adjustment
Automatic microprocessor controlled
V-Path Correction
Automatic microprocessor controlled
Measuring Range
.008″ to 25.00″ (.2 to 635 mm); depends
on material, probe, surface, condition, and
temperature
Digital Displayed Resolution
.001 or .01 inches; .01 or .1 mm
Reading Update Rate
4 or 8 Hz; 32 Hz in Min & B–Scan Modes
Gain
Automatic or adjustable in 1 dB steps
Receiver Bandwidth
500 KHz to 15 MHz
Material Velocity Range
39,400 to 393, 662 in/sec.; 1000 to
9999 m/sec.
Power Supply
4 ea. Standard, AA size, alkaline or nicad
rechargeable batteries
Fast Battery Charger
External, 5 hour maximum charge time
Battery Life (operating time)
40 hours continuous use at 4 Hz update
and backlight, Min Cap, and B–Scan off
Instrument Shut-off
Manual or automatic after 1 to 15 minutes
of non-use
I/O Port
Bidirectional RS232 selectable: baud rate
(maximum 115,200), printer, direct reports
Multilingual Operation
English, German, French, Italian, and
Spanish
Operating Temperature Range
+10° to +120°F (–10° to + 50°C)
Size
10.1″ H x 5.1″ W x 1.2″ D (256 mm x 129
mm x 30 mm)
Weight
1.6 lbs. (725 g) with batteries
Hazardous Area Operation
Safe operation as defined by MIL-STD810E, Method 511.3, Procedure 1.
Warranty
A full two-year warranty on parts and labor
Base Instrument Package
Operating Principle
Ultrasonic pulse/echo
Measurement Modes
Dual-element, Dual-element Dual-multi,
Single-element initial pulse-to-1st, singleelement multiple echo (peak and flank
modes), B-scan (optional TopCOAT & Auto-V
modes)
B-Scan
Cross-sectional view quickly identifies
thinned or laminated areas. Timed 3.5-21
seconds
DMS 2 A-Scan Ultrasonic Thickness Gauge
4 AA Alkaline Batteries
Rubber Field Case
Operating Manual
Certificate of Calibration
Firmware upgrade CD-ROM
PCCBL-841 9 Pin Serial PC Cable
PRTCBL-842 Parallel Printer Interface
Cable
PRTCBL-518 Serial Printer Cable
FB-143 Field Accessory belt with
removable pouches
LCD-124 Tearaway, protective screen
overlays
LCC-315 Locking Transport Case
UltraMATE® Windows™ Data
Management Software Program
AC-214 A/C Power Supply
Specifications
Memory Retention
5 years
File Structures
Linear, Grid, Boiler, Custom Point, Custom
Linear, Custom Grid, 3 or 4 level custom
and UltraPIPE files
Microgrids
Creatable on the job for all file structures
from 2 x 2 to 9 x 9 symmetrical sizes
Alphanumeric Input
On-command virtual keyboard at press of
direct access key
Inspection Comments/Notes
1 to 16 user-defined, 16 character
comments per reading
Store/Recall A-Scans and B–Scans Microgrids
Recall A-Scans for viewing, printing,
re-setting instrument or deleting
DMS2
Ultrasonic Thickness
Gauge
Accessories
Options
TopCOAT–018 Thru-Coating/Corrosion
Pitting/Coating Measurement
EM-166 Expanded Memory
Specifications
Data Recorder
Memory Capacity
Up to 150,000 thickness readings/1100
A-Scan Waveforms or B–Scan
crosssectional views with full instrument
parameters (expandable up to 318,000
thickness readings/2400 A-Scan
Waveforms or B–Scan cross–sectional views)
Reviews