Description
Fluke 9103-A-156 Dry Block Temperature Calibrator Offers
Features
Covers below-ambient temperatures as low as -25?C
Reaches temperatures 50?C below ambient, so -25?C is reached under normal ambient conditions
Stable to ?0.02?C
Display is calibrated to an accuracy of ?0.25?C at all temperatures within its range
In just 8 minutes, 0?C is reached, and 100?C is reached in 6 minutes
Choose one of four removable inserts sized for probes from 1/16″ to 1/2″ in diameter
These field dry blocks beat every other dry block in the industry in performance, size, weight, convenience, ease of calibration, software, and price
Heating and cooling rate of each of these dry blocks is adjustable from the front panel
Very easy to calibrate, you don’t even have to open the case; meaning less maintenance costs and less down time when they do need recalibration
Interface-it software lets you adjust set points and ramp rates, log dry block readings to a file, create an electronic strip chart, and perform thermal switch testing with data collection
Multiple hole inserts are available for a variety of probe sizes
Applications
Ideal for the following applications: Laboratory, Education, R&D, Calibration Labs, Field Calibration
Fluke 9103-A-156 Specifications
Temperature Range
-25 to 140?C?(-13 to 284?F) @ 23?C ?3?C ambient
Accuracy
?0.25?C, ?1?C in holes > 0.25″
Stability
?0.02?C @ -25?C; ?0.04?C @ -2140
Well-to-Well Uniformity
?0.1?C between similarly sized wells
Immersion Depth / Well Diameter
4.875″ (124mm)/ 1.25″ dia. x 4.88″ long (31.8 x 124mm)
Stabilization Time
7 minutes
Heating Time
18 minutes from ambient to 140?C
Cooling Time
20 minutes from ambient to -25?C
Click here for complete specifications on the Fluke 9103-A-156