Description
Features:
•Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors, transmitters and other instruments.
•Source/simulate volts, mA, thermocouples, RTDs, frequency, ohms, and pressure to calibrate transmitters.
•Power transmitters during test using loop supply with simultaneous mA measurement.
•Measure/source pressure using any of 29 Fluke 700Pxx Pressure Modules.
•Create and run automated as-found/as-left procedures to satisfy quality programs or regulations. Record and document results.
•Holds up to a full week of downloaded procedures and calibration results.
•Use many features like autostep, custom units, user entered values during test, one-point and two-point switch testing, square root DP flow testing, programmable measurement delay etc.
•Easy to use.
•Three year warranty.
•Bright white dual display. Read both sourced and measured parameters simultaneously.
•Multi lingual interface.
•Rechargeable NiMH battery for 10 hour uninterrupted use. Includes gas gauge.
•Handling of fast pulsed RTD transmitters and PLCs, with pulses as short as 1 ms.
•Delivered with DPC/Track Sample software.
•Compatible with many Asset Management software packages.
Included :
BC7240 battery charger
Li-on BP7240 battery pack
DPCTrack 2 Sample Software
Instruction manual
NIST-traceable calibration report and data
Three sets of TP220 test probes with three sets of extended tooth alligator clips
Two sets AC280 hook clips
C799 Soft Field Case
USB communication cable
Fluke 754HHC HART communications cable
Specifications:
Voltage DC
–/1 Year /2 Years:
100.000 mV/ 0.02%+0.005 mV /0.03%+0.005 mV
3.00000 V /0.02%+0.00005 V /0.03%+0.00005 V
30.0000 V /0.02%+0.0005 V /0.03%+0.0005 V
300.00 V /0.05%+0.05 V/ 0.07%+0.05 V
Voltage AC
Range 40 to 500 Hz /Resolution /1 Year/ 2 Years
3.000 V /0.001 V /0.5%+0.002 V /1.0%+0.004 V
30.00 V /0.01 V /0.5%+0.02 V /1.0%+0.04 V
300.0 V /0.1 V/ 0.5%+0.2 V /1.0%+0.2 V
Current DC
–/1 Year /2 Years :
30.000 mA /0.01% + 5 uA /0.015%+7 uA
110.00 mA /0.01% + 20 uA /0.015%+30 uA
Resistance
–/ 1 Year 2 /Years
10.000 O /0.05% + 50 mO /0.07%+70 mO
100.00 O /0.05% + 50 mO /0.07%+70 mO
1.0000 kO /0.05% + 500 mO /0.07%+0.5 O
10.000 kO /0.1% + 10 O/ 0.15%+15 O
Frequency
Resolution Accuracy (2 Years)
1.00 to 110.00 Hz /0.01 Hz 0.05 Hz
110.1 to 1100.0 Hz /0.1 Hz 0.5 Hz
1.101 to 11.000 kHz/ 0.001 kHz 0.005 kHz
11.01 to 50.00 kHz/ 0.01 kHz 0.05 kHz
Source Accuracy
Voltage DC
–/1 Year /2 Years
100.000 mV /0.01%+0.005 mV /0.015%+0.005 mV
1.00000 V /0.01%+0.00005 V/ 0.015%+0.0005 V
15.0000 V /0.01%+0.0005 V /0.015%+0.0005 V
Current DC
–/ 1 Year /2 Years
22.000 mA (Source)/ 0.01%+ 0.003 mA/ 0.02%+ 0.003 mA
Current Sink (Simulate) /0.02% + 0.007 mA/ 0.04% + 0.007 mA
Resistance
–/1 Year /2 Years
10.000 O /0.01% + 10 mO /0.015% + 15 mO
100.00 O /0.01% + 20 mO /0.015% + 30 mO
1.0000 kO /0.02% + 0.2 O /0.03% + 0.3 O
10.000 kO /0.02% + 3 O /0.03% + 5 O
Frequency
–/2 Years
0.1 to 10.99 Hz /0.01 Hz
0.01 to 10.99 Hz /0.01 Hz
11.00 to 109.99 Hz/ 0.1 Hz
110.0 to 1099.9 Hz/ 0.1 Hz
1.100 to 21.999 kHz /0.002 kHz
22.000 to 50.000 kHz/ 0.005 kHz
Technical Data
Data Log Functions
Measure functions: Voltage, current, resistance, frequency, temperature, pressure
Reading rate: 1, 2, 5, 10, 20, 30, or 60 readings/minute
Maximum record length: 8000 readings (7980 for 30 or 60 readings/minute)
Ramp Functions
Source functions: Voltage, current, resistance, frequency, temperature
Rate: 4 steps/second
Trip detect: Continuity or voltage (continuity detection not available when sourcing current)
Loop Power Function
Voltage: Selectable, 26 V
Accuracy: 10%, 18 V minimum at 22 mA
Maximum current: 25 mA, short circuit protected
Maximum input voltage: 50 V DC
Step Functions
Source Functions :Voltage, current, resistance, frequency, temperature
Manual Step :Selectable step, change with arrow buttons
Autostep :Fully programmable for function, start delay, stepvalue, time per step, repeat