Description
New! Fluke 754
You’re going to love what we’ve done with your calibrator.
The Fluke 754 is everything you needed and wished for in the 744 and more! HART communication capability is combined to deliver an integrated communicating calibrator. This rugged, reliable tool is ideal for calibrating, maintaining, and troubleshooting HART and other instrumentation.
The 754 does the work of several tools – sourcing, simulating and measuring pressure, temperature, and electrical signals in one rugged, hand-held device. You’ll fall in love with the new improved graphical screen, Li-on battery for longer life, USB port and new accessories to complete the package.
For documentation, the 754 automates calibration procedures and captures your data. And, of course, it helps you meet rigorous standards like ISO 9000, FDA, EPA and OSHA regulations. Supports popular instrumentation management software.
The 753 and 754, 743 and 744 work with the Fluke DPC/TRACK2? software, and with popular programs from Honeywell Meridium, Emerson, Cornerstone, Yokogawa, Prime Technologies, Intergraph and others. It allows you to create procedures, instructions, and action lists to deliver fast, easy documentation.
Fluke 754 Features
Work smarter. Work faster.
The 754 is a power multifunction documenting calibrator that lets you download procedures, lists, and instructions created with software-or upload data for printing, archiving, and analysis. The 754 also features a powerful built in HART? interface capable of performing nearly all the day-to-day tasks you now perform with a separate communicator.
Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors, transmitters and other instruments
Source/simulate volts, mA, thermocouples, RTDs, frequency, ohms, and pressure to calibrate transmitters
Power transmitters during test using loop supply with simultaneous mA measurement
Measure/source pressure using any of 29 Fluke 700Pxx Pressure Modules
Create and run automated as-found/as-left procedures to satisfy quality programs or regulations. Record and document results
Holds up to a full week of downloaded procedures and calibration results.
Use many features like autostep, custom units, user entered values during test, one-point and two-point switch testing, square root DP flow testing, programmable measurement delay etc.
Easy to use
Three year warranty
Bright white dual display. Read both sourced and measured parameters simultaneously.
Multi lingual interface
Rechargeable Li-Ion battery for 10 hour uninterrupted use. Includes gas gauge.
Handling of fast pulsed RTD transmitters and PLCs, with pulses as short as 1 ms.
Delivered with DPC/Track Sample software.
Compatible with many Asset Management software packages.
The Fluke 754 Documenting Process Calibrator-HART: Work smarter. Work faster.
The 753 and 754, 743 and 744 work with the Fluke software, and with popular programs from Honeywell Meridium, Emerson, Cornerstone, Yokogawa, Prime Technologies, Intergraph and others. It allows you to create procedures, instructions, and action lists to deliver fast, easy documentation
Fluke 754 Specifications
Range and 1 Year Accuracy
DC Voltage Measurement
100.00mV: 0.02% + 0.005mV
3.00000V: 0.02% + 0.00005V
30.0000V: 0.02% + 0.0005V
300.00V: 0.05% + 0.05V
AC Voltage Measurement
3.000V: 0.5% + 0.002V
30.00V: 0.5% + 0.02V
300.0V: 0.5% + 0.2V
DC Current Measurement
30.000mA: 0.01% + 5uA
110.00mA: 0.01% + 20uA
Resistance Measurement
10.000?: 0.05% + 50m?
100.00?: 0.05% + 50m?
1.0000k?: 0.05% + 0.5?
10.000k?: 0.1% + 10?
Frequency Measurement
1.00Hz to 110.00Hz: 0.05Hz
110.1Hz to 1100.0Hz: 0.5Hz
1.101kHz to 11.000kHz: 0.005kHz
11.01kHz to 50.00kHz: 0.05 kHz
DC Voltage Output
100.000mV:? 0.015% + 0.005mV
1.00000V: 0.015% + 0.0005V
15.0000V: 0.015% + 0.0005V
DC Current Output
22.000mA: 0.01% + 0.003mA
Current sint (simulate transmitter): 0.01% + 0.007mA
Resistance Sourcing
10.000?: 0.01% + 10m?
100.00?: 0.01% + 20m?
1.0000k?: 0.02% + 0.2?
10.000k?: 0.02% + 3?
Frequency Sourcing
Sinewave: 0.1Hz to 10.99Hz: 0.01 Hz
Squarewave: 0.01Hz to 10.99Hz: 0.01Hz
Sine and square 11.00Hz to 109.99Hz: 0.1Hz
Sine and square 110.0Hz to 1099.9Hz: 0.1Hz
Sine and square 1.100kHz to 21.999kHz: 0.002kHz
Sine and square 22.000kHz to 50.000kHz: 0.005kHz
Click here for complete specifications on the Fluke 754
Reviews
There are no reviews yet.