Description
Fluke 572-2 Offers
This high-temperature infrared thermometer is built for use in high-temperature industrial environments. It has a straight-forward user interface and soft-key menus, making even complex measurements easier. Quickly navigate and adjust emissivity, start data logging, or turn on and off alarms, with just a few pushes of a button.
Features
60:1 distance to spot ratio with dual laser sighting for fast, accurate targeting
Multi-language interface (user select)
Current Temperature plus MAX, MIN, DIF, AVG temperature displays
Compatible with standard mini-connector K-type thermocouples
Adjustable emissivity and predefined emissivity table
Infrared and thermocouple temperature on backlit display
Last reading Hold (20 seconds)
High and low temperature alarm
Data storage and review (99 data sets)
Tripod mount
12 or 24 hour clock
USB 2.0 computer interface cable
FlukeView? Forms Documenting Software
Two-year warranty
Applications
Utilities
Mining and Minerals
Oil and Gas
Fluke 572-2 Specifications
Infrared Measurements
Temperature Range
-22 to 1652?F (-30 to 900?C)
IR Accuracy (calibration geometry with ambient temperature 23?C ?2?C)
?32?F: ?2?F or ?1% of the reading, whichever is greater
?14?F to <32?F: ?4?F
<14?F: ?6?F
?0?C: ?1?C or ?1 % of the reading, whichever is greater
?-10?C to <0?C: ?2?C
<-10?C:?3 ?C
IR Repeatability
?0.5% of reading or ?1?F (?0.5?C), whichever is greater
Display Resolution
0.1?F/0.1?C
Distance:Spot
60:1 (calculated at 90% energy)
Minimum Spot Size
0.75″ (19 mm)
Laser Sighting
Offset dual laser, output <1 mW
Spectral Response
8 ?m to 14 ?m
Response Time (95%)
<500 ms
Emissivity
Digitally adjustable from 0.10 to 1.00 by 0.01 or via built-in table of common materials
Contact Measurements
K-Type Thermocouple Input Temperature Range
-454 to 2501?F (-270 to 1372?C)
K-Type Thermocouple Input Accuracy
<-40?F: ?(2?F + 0.2?/1?F)
?-40?F: ?1% or 2?F, whichever is greater
<-40?C: ?(1 ?C + 0.2?/1?C)
?-40?C: ?1% or 1?C, whichever is greater
K-Type Thermocouple
0.1?F/0.1?C
K-Type Thermocouple Repeatability
?0.5% of reading or ?1?F (?0.5?C), whichever is greater
Measurement Options
Hi/Low Alarms
Audible and two-color visual
Min/Max/Avg/Dif
Yes
Switchable Celsius and Fahrenheit
Yes
Backlight
Two levels, normal and extra bright for darker environments
Probe Input
K-type thermocouple simultaneous display of probe and IR temperature
Trigger Lock
Yes
Data Storage
99 points
Display
Dot matrix 98 x 96 pixels with function menus
Communication
USB 2.0
K-Type Thermocouple Specifications
Measurement Range (Bead Probe)
-40 to 500?F (-40 to 260?C)
Accuracy
?2.0?F (?1.1?C) from 32 to 500?F (0 to 260?C)
Typically within 2.0?F (1.1?C) from -40 to 32?F (-40 to 0?C)
Cable Length
40″ (1 m) K-type thermocouple cable with standard miniature thermocouple connector and bead termination
General Specifications
Safety and Compliance
IEC 60825-1
FDA Laser Class II
IEC 61326-1
CE Complaint
CMC # 01120009
Operating Temperature
32 to 122?F (0 to 50?C)
Storage Temperature
-4 to 140?F (-20 to 60?C)
Relative Humidity
10 to 90% RH non-condensing up to 86?F (30?C)
Operating Altitude
2000 meters above mean sea level
Weight
0.7099 lb (0.322 kg)
Power
2 x AA batteries
Battery Life
8 hours with laser and backlight on; 100 hours with laser and backlight off, at 100 % duty cycle (thermometer continuously on)
Click here for complete specifications on the Fluke 572-2
Reviews