Description
Fluke 3130-G2M/C Offers
The internal, motorized pump provides an easy, efficient alternative to time consuming, tiring hand pumps. For applications that involve filling a large volume with pressure, this instrument allows for connection to an external gas supply such as compressed plant air. The pressure can be fine-tuned using the variable volume.
Features
Measure and generate pressures from vacuum to 2 MPa (300 psi, 20 bar)
Internal pump can generate vacuum to -80 kPa (-12 psi, -0.8 bar) or pressure to 2 MPa (300 psi, 20 bar)
Supply pressure connection allowing the use of external gas supply up to 2 MPa (300 psi, 20 bar)
Includes variable volume for fine adjustment of pressures
Pressure measurement accuracy of 0.025% reading ?0.01% FS
Electrical measurement and 24 volt supply for close looped calibrations
Measure or generate 4 to 20 mA
Measure 0 to 30 V DC
Powered by internal, high capacity, NiMH battery or universal ac mains adapter
Compatible with the 700P Pressure Modules
Fluke 3130-G2M/C Specifications
Operating Temperature
-10 to 50?C
Storage Temperature
-20 to 60?C
Power Requirements
12 V DC
Power Source
Internal 3800 mAh advanced NiMH pack
Battery Life
Approx 50 hours (Measure only or external air; no pump)
EMI/RFI Conformance
EN61326:2006 Annex A
Connectors/Ports
1/8″ NPT
Dimensions
15.25 x 12 x 7″
Weight
7 kg (15 lbs)