Noyes C860 Quad Otdr And Certification Test Kit

$6,125

Category:

Description

Noyes C860 Quad OTDR and Certification Test Kit

The NOYES C860 QUAD Certification and OTDR Test Kit from AFL includes one hand-held C840 QUAD OLTS Tester and one C850 QUAD OTDR/OLTS with built-in auto test functionality. With this kit, technicians can troubleshoot and perform both Tier 1 and Tier 2 certification tests of MM and SM fiber networks, store results and create professional test reports.

The C850 is both a QUAD Certification Tester and full-featured QUAD OTDR in a compact case with a large transflective touch screen display suitable for both indoor and outdoor operation. The C850 features single-mode and multimode OTDR, Optical Light Sources (OLS), Visual Fault Locator (VFL, 650 nm), and an Optical Power Meter (OPM). As an OTDR, the C850 supports Full Auto, Expert (manual) and Real-Time test modes, simultaneous dual and single wavelength testing, and Event and Pass/Fail analysis based on default or user-defined thresholds. The C840 QUAD Certification Tester includes VFL, OPM, and both single-mode (1310/1550 nm) and multimode (850/1300 nm) OLS. The C840 may be used alone as a traditional power meter or light source to measure fiber loss or as a visual fault locator to find fiber breaks.

The C860 kit combines ease of use (Touch and Test™) with multiple functionality and supports visual inspection per IEC 61300-3-35 using the DFS1 Digital FiberScope allowing users the ability to view and document connector end-face images with their OTDR traces and loss results. Thousands of test results may be stored internally or on the supplied USB drive. Test results are transferable via a USB cable or USB drive to a computer for viewing, printing, and analyzing with the supplied Windows® compatible software – TRM® (Test Results Manager). Acceptance reports generated using TRM can include OTDR traces with summary and event information with or without pass/fail indication, Event maps, and end-face images.

Features:

  • OTDR dynamic range: 22 dB (MM); 26 dB (SM)
  • Inspection capable with the DFS1 Digital FiberScope
  • Integrated OPM, OLS, and VFL (650 nm)
  • OLS sources: LED – 850/1300 nm; Laser – 1310/1550 nm
  • Full Auto, Expert, Real-Time OTDR test modes
  • >8 hours battery life or AC power
  • Touch and Test™ user interface; TRM® reporting software
  • Automatic Pass/Fail analysis (TIA/ISO/EN)
  • Internal (1000s tests) and USB storage
  • Wave ID detect if used with NOYES Wave ID series light sources
  • Available in Seven languages

Applications:

  • Tier 1 and Tier 2 testing of premise networks
  • Bi-directionally measure loss and length of fiber links
  • Perform Pass/Fail Event and Link measurements using OTDR
  • Measure loss and length of fiber links
  • Certify SM and MM fibers using Pass/Fail criteria of industry standards, applications and user-defined thresholds
  • Create professional certification reports

Specifications:

OTDR MULTIMODE SINGLE-MODE
Emitter Type Laser
Safety Class Class I FDA 21 CFR 1040.10 and 1040.11, IEC 60825-1: 2007-03
Center Wavelengths 850/1300 nm 1310/1550 nm
Wavelength Tolerance ±20/30 nm ±20/30 nm
Dynamic Range (SNR = 1) 22 dB 26 dB
Event Dead Zone 1.5 m
Attenuation Dead Zone 9 m
Pulse Widths 10, 30, 100, 300 ns; 1, 3, 10 µs
Range Settings 250 m to 64 km 250 m to 208 km
Sampling Points Up to 16,000
Minimum Data Point Spacing 0.25 m
Group Index of Refraction (GIR) 1.4000 to 1.6000
Distance Uncertainty (m) ±(1 + 0.005 % x distance + data point spacing)
Linearity ±0.05 dB/dB
Loss Threshold 0.05 dB
Loss Resolution 0.01 dB
Reflectance Accuracy ±2 dB
VISUAL FAULT LOCATOR
Emitter Type Laser
Safety Class Class II FDA 21 CFR 1040.10 and 1040.11, IEC 60825-1: 2007-03
Wavelength 650 ±20 nm
Output Power (nominal) 0.8 mW
LIGHT SOURCE MM PORT SM PORT
Available Wavelengths (nom.) 850/1300 nm 1310/1550 nm
Emitter Type LED Laser
Safety Class Class I FDA 21 CFR 1040.10 and 1040.11, IEC 60825-1: 2007-03
Output Power >-20 dBm, 62.5 µm MM 0 dBm, 9 µm SM
Stability (after 15 minutes warm up) ±0.1 dB over 1 hour ±0.07 dB over 1 hour
±0.15 dB over 8 hours
Wave ID Transmit Yes
Tone Generation 270 Hz, 330 Hz, 1 KHz, 2 kHz
POWER METER
Calibrated Wavelengths 850, 1300 1310, 1490, 1550, 1625 nm
Detector Type InGaAs 2 mm
Measurement Range +6 to -60 dBm
Accuracy ±0.25
Measurement Units dB, dBm, mW
Wavelength ID Yes (to -47 dBm)
Set Reference Yes
Data Storage Yes
Tone Detection Yes (to -47 dBm)
GENERAL C850 OTDR C840 TESTER
Test Modes OTDR (Full Auto, Expert, Real-Time), Auto Test, OPM, OLS, VFL, DFS Auto Test, OPM, OLS, VFL, DFS
Trace File Format SR-4731 (GR-196-CORE Appendix A, B; SR-4731) N/A
Length Measurement Range 5 km (MM); 200 km (SM)
Data Storage Internal flash memory
USB flash drive (2.0)
Downloadable from unit directly to PC
Data Storage Capacity Internal >1000 fibers
Data Transfer to PC USB
Tool Free Adapters Modular cleanable SC/ST/LC
Size 27.4 x 19.3 x 7.1 cm (10.8 x 7.6 x 2.8 in) 23 x 11 x 7 cm (8.8 x 4.3 x 2.8 in)
Weight 2.3 kg ( 5 lb) 0.9 kg (2 lb)
Operating Temperature -10 °C to +50 °C, 0 to 90 % RH (non-condensing)
Storage Temperature -20 °C to +60 °C, 0 to 90 % RH (non-condensing)
Power Rechargeable Li-Ion or AC power adapter
Battery Life >8 hours continuous testing
Recharge Time 4 hours
Display 16.51 cm (6.5 in), color, transflective 9.65 cm (3.8 in), color, transflective

Reviews

There are no reviews yet.

Be the first to review “Noyes C860 Quad Otdr And Certification Test Kit”

Your email address will not be published. Required fields are marked *