Description
Fluke 9100S-A-156 Dry Block Calibrator Offers
Features
Temperature range to 375?C (707?F) is perfect for checking RTDs, thermocouples, and small bimetal thermometers in the field
With a proprietary Hart Scientific temperature controller, it has a display resolution of 0.1?, an accuracy of ?0.5?C, a stability of ?0.1?C, and a stabilization time of just five minutes
Simple and convenient, plug it in, switch it on, set the temperature with the front-panel buttons, and insert your probe into the properly sized well
Compare the reading of your device to the display temperature or to an external reference, and the difference is the error in your device
Despite its small size (2-1/4″ high and 5″ wide) and light weight, it out-performs every dry-well in its class
Block “A” handles a full range of probe sizes with a single well of each size (1/16″, 1/8″, 1/4″, 5/32″; 2 wells of 3/16″)
Applications
Ideal for the following applications: Laboratory, Education, R&D, Calibration Labs, Field Calibration
Fluke 9100S-A-156 Specifications
Temperature Range
35 to 375?C?(95 to 707?F)
Accuracy
?0.25?C @ 50?C; ?0.25?C @ 100?C; ?0.5?C @ 375?C
Stability
?0.07?C @ 50?C; ?0.1?C @ 100?C; ?0.3?C @ 375?C
Well-to-Well Uniformity
?0.2?C with sensors of similar size at equal depths within wells
Well Depth
4″ (102mm); 1/16″ (1.6mm) hole is 3.5″ (89mm) deep
Stabilization Time
5 minutes
Heating Time
Ambient to 375?C: 9.5 minutes
Cooling Time
375?C to 100?C: 14 minutes
Click here for complete specifications on the Fluke 9100S-A-156
Reviews
There are no reviews yet.