Description
Fluke 1750-TF/ET Offers
New Fluke 1750 Windows 10 Tablet: New Window’s 10 Tablet for Control, Setup, Data Download and Analysis with Fluke 1750 in the field now included.
This new tablet runs the full version of Fluke Energy Analyzer for instrument setup in the field and downloading with basic analysis. Setup and download is made by connecting the 1750 to the tablet via a wired Ethernet connection. The tablet has a detachable keyboard with two USB ports. The supplied Ethernet adapter connects to the other USB ports and then to the instrument via a standard Ethernet cable.
Power quality that meets the standard: All measurements comply with IEC61000-4-30 standards for correct evaluation of all measured values including voltage, current, power, harmonics, flicker etc.
Features
Quick and reliable configuration: ARCHOS 43 Internet Tablet provides a window into what the instrument is recording, enabling quick and reliable configuration even in awkward test locations
Threshold-free set up: Apply thresholds after data is collected with Fluke Power Analyze Software ? no need to worry about missed information due to incorrect setups
Captures everything: Cross-channel and current triggering capture every measurement, on every channel, every time
Intuitive PC software: Easily analyze data and generate reports with automated EN50160 reporting and compliance
Reporting has never been easier: Auto Report creates either standard reports or customized reports with the minimum effort or hassle.
Plug and play: Set up in minutes with self-identifying current probes and single-lead voltage connections
No need to reconnect wires: Swap channels internally with the wireless PDA or PC when connections are not correct
Measure every parameter: Voltage and current on three phases, neutral, and ground
5 MHz, 8000 Vpk waveform capture: Get a detailed picture of even the shortest events
Quickly retrieve data: With included SD memory card or via the 100BaseT high-speed Ethernet connection. SD card includes automatic download to SD when no other connection is made to the instrument.
Applications
Long-term analysis: Uncover hard-to-find or intermittent issues; monitor critical equipment, capturing power quality events to correlate with equipment malfunction
Power quality surveys: Quantify power quality throughout a facility, documenting results with professional reports
Quality of service compliance: Validate incoming power quality at the service entrance
Equipment Installation/Commissioning: Benchmark: power system prior to install to insure quality of service
Long-Term Analysis ? Uncover hard-to-find or intermittent issues
Load Studies ? Verify the available electrical system capacity before adding loads
Fluke 1750-TF/ET Specifications
Input channels
Voltage: 4 channels, AC/DC / Current: 5 channels
Voltage channels
Input resistance: 2 M? / Input capacitance: < 20 pF
Measurement Type
True rms calculated continuously: every cycle, and every 10 or 12 cycles at 50 or 60 Hz respectively, as required by standards
Frequency Measurement Range
42.5 Hz to 69 Hz
Power Factor Measurement Range
0.000 to 1.000
Connector
RJ-45
Speed and Type
10/100 Base-T, auto MDIX
Communications Protocol
TCP/IP over ethernet
Voltage Terminals
1100 V rms
Enclosure Protection
IP30 (per EN 60529)
Click here for complete specifications on the Fluke FLUKE-1750-TF/ET
Reviews
There are no reviews yet.